Chapter 12
| System control
|
Dionex IC Pure system
74

Set the interval pump time

The interval pump time is the amount of time the pump is working to recirculate water in the system.
The standard setting is 5 minutes of pump recirculation for every 30 minutes that the system stands
still during "Interval mode". The majority of systems do not need this setting to be changed.
Basic setting: 5 minutes
Setting range: 1 - 30 min.
After entering the OEM menu, press the "Menu" button four times. The display shows:
Note The system is factory adjusted in the system control. In this section it is not necessary to be
carried out changes of value for operating with an IC system.

Set the interval rinse time

In this menu the rinse interval time can be set. When the system operates in the Interval mode, the
system will rinse water to drain for 0.5 seconds every 30 minutes.
Basic setting: 0.5 sec.
Setting range: 0.1 - 2 sec.
After entering the OEM menu, press the "Menu" button five times. The display shows:

Set the real-time clock

The real-time clock can be set in this menu.
Basic setting: The actual date
Setting range: 1 - 12 month; 1 - 31 day; 0 - 24 h; 0 - 60 min.
After entering the OEM menu, press the "Menu" button six times. The display shows: