Thermo Scientific Barnstead Pacific RO 3 Operating Instructions Manual

Water purification system

Advertisement

Quick Links

Operating Instructions
Thermo Scientific Barnstead Pacific RO
Water Purification System
 
Art. no.: 50132385 RO 3
 
Art. no.: 50132386 RO 7
 
Art. no.: 50132387 RO 12
Serial no.: ..........................
Read these Operating Instructions prior to system installation and start-up!
50132707, State 11.12 Rights to technical changes reserved!
 
Art. no.: 50132388 RO 20
 
Art. no.: 50132389 RO 40

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific Barnstead Pacific RO 3

  • Page 1 Operating Instructions Thermo Scientific Barnstead Pacific RO Water Purification System     Art. no.: 50132385 RO 3 Art. no.: 50132388 RO 20     Art. no.: 50132386 RO 7 Art. no.: 50132389 RO 40  ...
  • Page 2 Thermo Scientific Barnstead Pacific RO Pure water system EC-Declaration of Conformity in accordance with the EEC machine directive 2006/42/EC, appendix II A We hereby certify that the following described machine in its conception and form put by us into circulation is in accordance with all the relevant essential health and safety requirements of the EC machinery directive 2006/42/EC as amended and the national laws and regulations adopting this directive.
  • Page 3: Preface

    Thermo Scientific Barnstead Pacific RO Pure water system Preface Dear Sir or Madam, In deciding to purchase a pure water system of type Pacific RO you have selected a high-quality product. Thank you for the confidence you have placed in us.
  • Page 4: Table Of Contents

    Thermo Scientific Barnstead Pacific RO Pure water system 1. Contents Preface ......................3   1. Contents ....................4   2. Explanatory notes on the operating instructions ........6   3. Transport and packaging ................. 7   3.1 Examination on receipt ..................7  ...
  • Page 5 Thermo Scientific Barnstead Pacific RO Pure water system   12.2.3 Recirculation time: ..................... 29  12.2.4 Rinsing time: ......................29  12.2.5 Interval rinse time: ..................... 30  12.2.6 Real-time clock: ......................30  12.2 7 Sending interval: ....................... 30  12.2.8 Languages: ....................... 31 ...
  • Page 6: Explanatory Notes On The Operating Instructions

    Thermo Scientific Barnstead Pacific RO Pure water system 2. Explanatory notes on the operating instructions EU Mark of Conformity CSA - admission Important operating and/or maintenance instructions! Read the operat- ing instructions with due care . Risk of electric shock! Electrical work on the system is only to be car- ried out by qualified personnel.
  • Page 7: Transport And Packaging

    Thermo Scientific Barnstead Pacific RO Pure water system 3. Transport and packaging Pure water systems are carefully checked and packed prior to shipment, but there is neverthe- less always a possibility that damage could occur to them during shipment. 3.1 Examination on receipt ...
  • Page 8: Safety Precautions

    Thermo Scientific Barnstead Pacific RO Pure water system 4. Safety precautions For your own safety, please observe the above safety precautions! Your Pacific RO system is a modularly constructed, pure water system that  serves exclusively for the purification of tap water.
  • Page 9: Intended Use

    Thermo Scientific Barnstead Pacific RO Pure water system 5. Intended use The Pacific RO pure water system is a reaction to the continually increasing requirements that water of pure quality must fulfil, the increasingly strict demands resulting from technological ad- vances and the need for user-friendly systems and complete solutions.
  • Page 10: Extent Of Delivery

    Thermo Scientific Barnstead Pacific RO Pure water system 6. Extent of delivery The Pacific RO pure water system consists of: Pacific RO Article no.: 5013xxx__ Assembly kit (RO 3-20) Article no.: 50132708 Assembly kit (RO 40) Article no.: 50134364 consisting of: 2 x PVC connecting hose, 1.5 m, straight/angled...
  • Page 11: Technical Specifications

    Thermo Scientific Barnstead Pacific RO Pure water system 7. Technical specifications Demands on the feedwater Source Potable tap water that has been softened or hardness stabilized. Silt density index (SDI) < 5, With higher values, a pretreatment (article no. 09.4000) must be installed upstream of the system Resistance >0.001 MΩxcm...
  • Page 12 Thermo Scientific Barnstead Pacific RO Pure water system Electrical connections / external switched mode power supply RO 3-20 Input voltage AC 100 – 240 V, 50 – 60 Hz, 5 – 3.8 A Output voltage DC 24 V, 3.8 A...
  • Page 13: Flow Chart

    Thermo Scientific Barnstead Pacific RO Pure water system 8. Flow chart 8.1 Flow chart, tank without pump 8.2 Flow chart, tank with pump...
  • Page 14 Thermo Scientific Barnstead Pacific RO Pure water system B1 Storage tank: Stores the pure water that is produced. F1 5µm Prefilter + activated Prevents the entry of particles > 5µm carbon and guards against too high free chlorine concentrations. F2 Hardnes stabilization Binds calcium and so prevents precipitation of it.
  • Page 15: How The System Functions

    Thermo Scientific Barnstead Pacific RO Pure water system 8.3 How the system functions After switching the ON/OFF-key to on, the system starts either in the production mode or stand- by mode, according to the filling level in the storage tank.
  • Page 16 Thermo Scientific Barnstead Pacific RO Pure water system Your pure water system is equipped with automatic flushing. Flushing is carried out when the system is switched on, at the end of each production, and also every 12 hours. For this, rinsing...
  • Page 17: The Installation Area

    Thermo Scientific Barnstead Pacific RO Pure water system 9. The installation area The following criteria must be taken into consideration when choosing the installation area.  Minimum temperature in the installation area: + 2°C - +40°C.  The surface that your pure water system is to be stood on must be strong enough to support it (for weight, see "Technical specifications").
  • Page 18: Putting The System Into Operation

    Thermo Scientific Barnstead Pacific RO Pure water system 10. Putting the system into operation TKA-Pacific NONSTOP enter menu INTERVAL 1. Connect the raw water inlet of your pure water system to the pretreatment and this to a water tap that can be shut off.
  • Page 19 Thermo Scientific Barnstead Pacific RO Pure water system Check the connections for perfect leakproof connection. Check the raw water pressure. It must be within the permissible pressure range (see the Technical Specifications). Open the raw water tap. Before you now switch the system on, please read through the rinsing procedure for removing preservative solution from reverse osmosis membranes supplied filled with it in the „Cleaning and disinfecting“...
  • Page 20: Wall Mounting

    Thermo Scientific Barnstead Pacific RO Pure water system 10.1 Wall mounting Proceed as follows to mount your Pacific RO system to a wall: 1) Use a twist drill (8mm or 5/16 inch) to make the two holes in the wall that are required as shown in the diagram above.
  • Page 21: Mounting The Power Pack (Voltage Supply)

    Thermo Scientific Barnstead Pacific RO Pure water system 10.2 Mounting the power pack (voltage supply)  Whenever possible, mount the power pack on the wall to the left or right of the pure wa- ter system where it is freely accessible.
  • Page 22: Operating Elements

    Thermo Scientific Barnstead Pacific RO Pure water system 11. Operating elements Stand by 0.22 MΩxcm Schaltet das System an und aus Erhöht im Menu die Zahlenwerte Bestätigt in den Menüpunkten den Eingabewert Ruft die Menüs auf und schaltet in nächsten Menüpunkt...
  • Page 23: System Control

    Thermo Scientific Barnstead Pacific RO Pure water system 12. System control General information When the ON/OFF key is pressed, the pure water system starts running either in the operating mode or in the stand-by mode, as governed by the float switch.
  • Page 24: Permeate Limiting Value

    Thermo Scientific Barnstead Pacific RO Pure water system 12.1.2 Permeate limiting value: A second press on the menu-key allows the permeate limiting value to be set. Should this limiting value be exceeded, then the Lim.val.permeate message is shown (measuring position LF 2).
  • Page 25: Pretreatment Operating Hours

    Thermo Scientific Barnstead Pacific RO Pure water system 12.1.4 Pretreatment operating hours: A fourth press on the menu-key in this menu brings the operating hours of the pretreatment car- tridge to display. This operating time has a limiting value, which is set in the UV menu. The fault message that is displayed when the limiting value is exceeded is "Pretreatment".
  • Page 26: Disinfection

    Thermo Scientific Barnstead Pacific RO Pure water system 12.1.6 Disinfection: (Is not active in this version) The display shows: Disinfection Press Enter ! 12.1.7 Fault storage: A seventh press on the menu-key in this menu calls the fault storage prompt. Confirmation of this with the Enter-key allows the fault storage to be looked through.
  • Page 27: Unlocking The System

    Thermo Scientific Barnstead Pacific RO Pure water system 12.1.8 Unlocking the system: An eighth press on the menu-key in this menu brings you to the “Code” menu. To prevent unau- thorized access to the settings in the system control, changes to the settings can only be carried out when a correct code from the Assignment Table that follows is entered and confirmed with the Enter-key.
  • Page 28: Oem Menu

    Thermo Scientific Barnstead Pacific RO Pure water system 12.2 OEM Menu Basic settings and limiting values can be changed in this menu. To make changes in the OEM menu, the system control must previously be unlocked (see 11.1.8). Calling the OEM menu: Simultaneous pressing of the Interval-key and the Nonstop-key calls the OEM menu.
  • Page 29: Disinfection Time

    Thermo Scientific Barnstead Pacific RO Pure water system 12.2.2 Disinfection time: A second press on the menu-key: (this is not active in this version) The display shows: OEM menu Disinfect. time 30 min. 12.2.3 Recirculation time: A third press on the menu-key: (this is not active in this version) The recirculation time is set in this menu.
  • Page 30: Interval Rinse Time

    Thermo Scientific Barnstead Pacific RO Pure water system 12.2.5 Interval rinse time: A fifth press on the menu-key: The interval rinse time can be set in this menu. Rinsing is then carried out for this time period when the operating mode is changed, between stand-by and operation and every 12 hours.
  • Page 31: Languages

    Thermo Scientific Barnstead Pacific RO Pure water system 12.2.8 Languages: An eighth press on the menu-key The language in which the texts are displayed is set in this menu. The choice is of English, German or French. Basic setting: English...
  • Page 32: Adjusting The Float Switch Circuit Hysteresis

    Thermo Scientific Barnstead Pacific RO Pure water system 12.2.11 Adjusting the float switch circuit hysteresis: An eleventh press on the menu-key: Basic setting: Off: 100 % 70 % Setting range: Off: 25 - 100 % 0 - 70 % With a setting over 100 % for the upper switching point, the display of the tank level is switched off, so that this setting shows whether an analogue or a digital float switch is installed.
  • Page 33: Printer Output

    Thermo Scientific Barnstead Pacific RO Pure water system 12.3 Printer output Various parameters can be recorded by means of a printer. A distinction is made between three types of messages: - Standard messages - Code messages - Fault messages 12.3.1 Standard messages: A record of all measured values is printed out here in dependence on the sender interval.
  • Page 34: Maintenance

    Thermo Scientific Barnstead Pacific RO Pure water system 13. Maintenance Your system requires regular, proper and professional maintenance. We recommend that you secure a service contract to ensure that the necessary maintenance work is carried out. You then have the certainty of a high operational safety and reliability.
  • Page 35: Maintenance Intervals

    Thermo Scientific Barnstead Pacific RO Pure water system 13.1 Maintenance intervals Consumable materials are to be replaced at the intervals given in the following Table or when there is a drop in performance: Material Flow chart no. Article no. Interval* Pretreatment 09.4001...
  • Page 36: Replacing The Ro Membrane

    Thermo Scientific Barnstead Pacific RO Pure water system 13.3 Replacing the RO membrane Pacific RO 3 – 12: 1 RO membrane Pacific RO 20 – 40: 2 RO membranes Ansicht von hinten - ohne Rückwand Back view, with back panel removed ...
  • Page 37: Changing The Filter Cartridge

    Thermo Scientific Barnstead Pacific RO Pure water system 13.4 Changing the filter cartridge For changing the filter cartridge,proceed as follows: 1. Switch off your device and turn off the feedwater supply. 2. Press pressure knob (1) to unlock and remove cover (2).
  • Page 38: Disinfection

    Thermo Scientific Barnstead Pacific RO Pure water system 13.5 Disinfection Your system should be cleaned and disinfected at least once a year to eliminate any bacteria that are possibly in the system. We recommend that you carry out cleaning and disinfection shortly before the time that the filter cartridge must to be replaced.
  • Page 39 Thermo Scientific Barnstead Pacific RO Pure water system Disinfection of the tank/ recirculation Switch the Pacific system off. Shut off the supply of feedwater to the system and release pressure from the feed line. Half empty the tank belonging to the system, screw the lid off and pour the contents of a box respectively a syringe of the cleaning solution into it.
  • Page 40: Waste Disposal

    Thermo Scientific Barnstead Pacific RO Pure water system 14. Waste disposal When the packaging is no longer needed it can be disposed of as household waste. Systems are in conformity with EEC Guideline 2002/95/EC The system is not to be thrown away as household waste but must be properly disposed of. It can be returned to the manufacturer for safe disposal according to EEC Guideline 2002/96/EC.
  • Page 41: Trouble Shooting

    Thermo Scientific Barnstead Pacific RO Pure water system 15. Trouble shooting Fault Cause Remedy The system does not start - No supply of power - Supply power No dispensing possible - Feedwater tap is closed - Open the feedwater tap...
  • Page 42 “Measuring cell LF3” - System control is defect - Replace the system control The address to contact when your system requires service: Overview of Thermo Scientific International Sales Organization Postal address USA: Thermo Scientific 275 Aiken Road Asheville, NC 28804...
  • Page 43 Thermo Scientific Barnstead Pacific RO Pure water system Enquiries from Europe: Austria Sales: +43 1 801 40 0 Service: +43 1 801 40 0 Belgium Sales: +32 53 73 4241 Service: +32 53 73 4241 Finland/Nordic/Baltic countries Sales: +358 9 329 100...
  • Page 44: Replacement Parts Pacific Ro 3 - 40

    Thermo Scientific Barnstead Pacific RO Pure water system 16. Replacement parts Pacific RO 3 – 40 Pos. R+I no. Article designation Article no. Pressure booster pump, RO (Pacific RO 3 – 40) 19.0050* RO Module, consisting of: Reverse osmosis membrane (Pacific RO 3 – 12) 22.0046*...
  • Page 45: Replacement Parts Tank (Optinal)

    Thermo Scientific Barnstead Pacific RO Pure water system 17. Replacement parts tank (optinal) Pos. R+I no. Article designation Article no. Pure water tank, 30 litre 18.0114 Pure water tank, 60 litre 18.0115 Dispensing valve 14.0250 LIS 100 Float switch for 30 litre tank 16.0303...
  • Page 46: Consumable Materials And Accessories

    Thermo Scientific Barnstead Pacific RO Pure water system 18. Consumable materials and accessories Pacific RO Article Article designation 09.4000 Pretreatment for RO Complete, incl. 2 x 10“ housings, hardness stabilization and Activated carbon combi-cartridge 06.5201 Combi-cartridge with activated carbon, 10“...
  • Page 47: Terminal Assignment

    Thermo Scientific Barnstead Pacific RO Pure water system 19. Terminal assignment 19.1 Pacific RO 3-20 (24V)
  • Page 48: Pacific Ro 40 (48V)

    Thermo Scientific Barnstead Pacific RO Pure water system 19.2 Pacific RO 40 (48V)
  • Page 49: Maintenance Record

    Thermo Scientific Barnstead Pacific RO Pure water system 20. Maintenance record (Please note that correct keeping of this record is a condition of the guarantee) Customer address: Location: System type: Serial no.: Year made: Date Resistance, Resistance, Prefilter Hardness Signature...