Thermo Scientific dionex ic pure water Operating Instructions Manual page 95

Table of Contents

Advertisement

Step
Action
5
a. After 30 minutes, disinfection cycle completely drains the
Note To drain the content of the 5 L tank, the Dionex IC Pure
system must be in "Nonstop" mode.
b. If the external 20 L tank is also connected, switch the unit off
c. Attach a drain fitting to the supply tube and drain the
d. After this, run one complete tank filling of feedwater through
e. Switch the Dionex IC Pure system off and refill the tank
Note If the external 20 L tank with level control is connected it
will be filled automatically. If the instrument is not attached to a
feedwater supply, fill the tank manually.
f.
Note To dispense the amount of 10 L of water, the Dionex IC
Pure system must be in the Nonstop mode.
Thermo Scientific
tank content through the sampling tap. To completely rinse
out the disinfection solution and contaminated water, the
tank is filled up and drained twice.
and remove the supply line from the connection on the
Dionex IC Pure system for the 20 L tank.
complete tank.
the 20 L tank and then reattach the supply line for the 20 L
tank to the connection on the Dionex IC Pure system.
completely with feedwater.
Switch the unit back on and dispense around 10 L of water
from the external tank at the sampling tap.
Figure
b.)
o.d. 8 mm (0.31 in) tubes
c.)
Supply tube from external 20 L
tank to the unit
Drain fitting
Maintenance |
Chapter 13
Dionex IC Pure system | 85