Chapter 13
| Maintenance
Step
Action
3
Always wear protective gloves and goggles when handling the
disinfecting agent cleaning solution.
a. Fill the internal 5 L tank or the external 20 L tank 80% with
b. Switch on the unit, open the sampling tap, extract 1 L of
Note There is no cause of worry should you notice a chlorine
smell during the entire disinfection process, as there is no risk
of exceeding the limit for chlorine gas in closed rooms.
4
Keep pressing the "Menu" button until "Code" appears in the dis-
play and then press "Enter".
Note For this action you must have a Level 1 code. Refer to
section "Entering a Code number" on page 66 for the code to be
used for this and how to enter the code.
a. In the user menu, press the "Menu" button to scroll to
b. The display shows "Disinfection cartridge". Press "Enter" to
c. The disinfection process is then started.
Note Disinfection is completed after 30 minutes. The
remaining time is displayed.
|
Dionex IC Pure system
84
water and pour the content of one syringe cleaning solution
into the tank. You can view the level of both tanks at the
additional tank indicator on the unit.
water and then close the sampling tap.
"Disinfection". Press "Enter".
confirm that Disinfection cartridge or Disinfection adapter
has been installed.
Figure
a.)
b.)
c.)