Chapter 13
| Maintenance
Step
Action
1
Switch the Dionex IC Pure system off.
2
Remove the cover for the ultrapure cartridge.
3
a. Take the ultrapure cartridge out of the unit (see "Changing the
b. Use the Phillips screwdriver to unscrew the two screws from
4
Note To easily remove the UV lamp in the next step, pull the
UV lamp out by about 1 cm while still plugged in.
5
a. Carefully unplug the plug for the UV lamp.
b. Now, carefully pull the UV lamp up while turning it slightly in
|
Dionex IC Pure system
90
Always wear safety gloves when changing the UV-lamp, in order to prevent that your skin comes in
contact with the UV-lamp glass.
Wear directly a breathing protector, filter category FFP3, when you are seeing that the glass of the
UV-lamp is broken and ventilate the room well.
Note Contact your local service organization for proper disposal of the broken UV-lamp.
Note The Hg content in the UV-lamp is so low so that no damage to the environment can arise.
ultrapure cartridge" on page 79). If an optional pretreatment
cartridge is installed, remove it too (see "Replacing the
pretreament cartridge (optional)" on page 80).
the bracket for the UV assembly.
a clockwise direction at the same time.
Figure
b.)
UV lamp
Mounting
plate
UV-lamp
UV- lamp
casing
Screws