100
Error
Dispensed water flow rate is too low
Display of TOC content permanently
at 99 ppb
No analyzer is indicated at the
Dionex IC Pure system for operation
with an analyzer.
Wrong time or date
Wrong language
Error message: "Limit value feed"
Display indicates +IN
Error message:"Pur.limit"
Error message:"UV-time"
Cause
UF-Module blocked
Initial pressure too low (external
20 L tank with level control)
Internal pressure too low
Dispensed water flow rate out of
the tolerance range
0.2 µm final filter is clogged
Air look in unit
UV lamp not burned in / counter
reset
TOC measuring cell defective
Settings in OEM menu are
incorrect
IC system not activated
Cable to IC system not
connected
Cable break
Time zone
Summer/winter time
Wrong language set
Feedwater conductivity too high
Limiting value set too low
Measuring cell cable break
Ultrapure cartridge exhausted
Limiting value set too low
UV lamp operating time has
been exceeded
Remedy
Replace UF-module (see
"Changing the ultrafilter" on
page 86)
Set the pressure regulator
Readjust pressure reducer
Contact Thermo Scientific for
calibration
Replace the 0.2 µm final filter
Dispense water at POU or rinse to
purge air from the system
Replace the UV lamp and reset the
operating hours counter (see
"Replacing the UV lamp" on
page 89)
Replace the TOC measuring cell
Correct the settings in the OEM
menu (see "OEM menu" on
page 72)
Activate IC system
Connect the cable to the IC system
Replace the cable
Reset time and date
Correct language setting
Check the feedwater measuring
cell and deactivate in the menu
Check and reset the limiting value
Replace measuring cell
Replace ultrapure cartridge (see
"Changing the ultrapure cartridge"
on page 75
Check and reset the limiting value
Replace the UV lamp (see
"Replacing the UV lamp" on
page 89)
Reset the operating time counter
Thermo Scientific