Chapter 15 Troubleshooting - Thermo Scientific dionex ic pure water Operating Instructions Manual

Table of Contents

Advertisement

Error
The system does not start
Dispensing not possible
Resistance < 18.2 M·cm
Control panel non responsive
Water leak
Thermo Scientific
Troubleshooting
Note Contact the service department if you cannot rectify this error.
Cause
No supply of power
Feedwater supply line for
external tank filling station
closed
Feedwater and rinse water
connections are mixed up
Feedwater pressure < 100 kPa
(14.5 psi)
Sterile filter blocked
Pretreatment cartridge blocked
Cartridge is exhausted
Temperature compensation
deactivated, recalibration
required
Microprocessor locked up
PCB error
Leaky tubing connection
Feedwater pressure > 600 kPa
(87 psi) (external tank filling
station only)
Remedy
Provide power
Open the feedwater supply line
Replace connections
Increase feedwater pressure
Replace the sterile filter (see
"Replacing the 0.2 µm final filter"
on page 92
Install new pretreatment cartridge
Check feedwater quality
Activating temperature
compensation ('TC' should appear
in the display) Contact the service
organization for calibration
Unplug the mains plug for 5
seconds.
Contact the Thermo Fisher
Scientific service department
Check and seal the tubing
connection
Install a pressure reducer
Contact the Thermo Fisher
Scientific service department
Dionex IC Pure system | 99