Rail Mounting; Bolt Specifications; Select Rails; Boom Mounted For Monitor Suspension - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
For additional details on ceiling requirements for stationary rails, refer to: - Direction 46–019639,
Advantx (VHLA) XT Stationary Rails Installation and Adjustment .
When evaluating ceiling you must take into account the following mounting information:

3.4.1 Rail Mounting

Attach stationary rails to structural steel with through-bolts in concrete ceilings. Do not use screw
anchors in direct tension.
Mount stationary rails directly to the ceiling slab or to flush–mounted unistrut or halfen structure.
In higher rooms with false ceiling, mount stationary rails to rigid vertical members hung from ceiling
slab.
Securing a supplementary channel to the bottom of the vertical members and mounting the
stationary rails to this channel can greatly reduce the number of vertical members.
The stationary rail support structure must be leveled before installation can begin. Do not assume
that any support structure is level within specified tolerances, particularly after removing
suspensions from an existing room.

3.4.2 Bolt Specifications

The maximum load per bolt will not exceed 1557 N.
Each bolt must not "pull out" or otherwise fail under a vertically downward
N.

3.4.3 Select Rails

All XT Stationary rails are with a select length process. Detail of available length is illustrated in
Chapter 7, Section 4, European Process Order

3.4.4 Boom Mounted for Monitor Suspension

One type of boom mounted is supplied as counterpoised monitor suspension not adjusted on site
as follows:
- for 2 x 21" (53 cm) EIZO GmbH (formerly Siemens) flat panel 21" (Hi–bright) and additional AW
NEC monitor.
Chapter 3 Physical Requirements of Room
Direction 5160944-13-1EN, Revision 2
Select.
dead load of 6227
75

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents