Viewing an Intensity Scan of the Deuterium Lamp

Thermo Scientific
To view an intensity scan of the deuterium (D
1. Create a display method to view the intensity of the deuterium lamp. Set channels A,B,
and C to monitor diodes 35, 52, and 77, respectively (see
See
"Creating a Display Method to View the Light Intensity"
on creating a display method.
Figure 48. Surveyor PDA Plus Method page with the selection of the Diodes/Intensity option
2. Load the display method to the detector as follows:
a. From the menu bar, choose Surveyor PDA Plus > Direct Control to open the
Direct Control dialog box.
b. Click the Display tab to open the Display page.
c. In the Control area, click Load Method to open the Open Display Method
dialog box.
d. Browse to the display method that you created to monitor the deuterium lamp. Then
click Open to load the method.
The filename of the display method (.spda) appears in the box above the Spectrum
plot.
3. Click the Configuration tab to open the Configuration page. Then turn off the tungsten
lamp.
4
Xcalibur Diagnostics for the PDA Detector
Verifying Operational Performance
) lamp
2
Figure
48).
on
page 45
Surveyor PDA Plus Detector Hardware Manual
for instructions