Omron ZW Series User Manual page 24

Displacement sensor, confocal fiber type
Hide thumbs Also See for ZW Series:
Table of Contents

Advertisement

Introduction
CONTENTS
I/O setting
measurement
[I/O]
[HLD.RST]
Judgment
Analog output
[ANALOG]
*
The same output destination is set for all banks. The output destination cannot be set separately for individual banks.
ZW
22
User's Manual
Non
setting
Output object
[JUDGE]
[OUTPUT]
Hysteresis
[HYS]
Timer
[TIMER]
Voltage or
current
*
[V OR C]
Output object
[OUTPUT]
Focus
[FOCUS]
Calibration
[CALIB]
Setting
Option/Setting range
Non
measurement
Keep [KEEP]/Clamp [CLAMP]
setting
[HLD.RST]
Current output: MIN [MIN]
(Approx. 3.4 mA)/4 mA/5 mA/.../19
mA/20 mA/MAX [MAX]
Analog
(Approx. 21 mA) (Default: MAX)
output
Voltage output: MIN [MIN]
[ANALOG]
(Approx. -10.8 V)/-10 V/-9 V/.../9 V/
10 V/MAX [MAX] (Approx. 10.8 V)
(Default: MAX)
Binary
output
MAX [MAX]
*
/MIN [MIN]
[BINARY]
Output object
TASK1 [TASK1]
[OUTPUT]
TASK3 [TASK3]/TASK4 [TASK4]
0 to 99.9999 mm
Hysteresis
(Default: 0.05 % of measurement
[HYS]
range)
Timer
Off
*
[OFF]/Off delay [OFF.DLY]/On
[TIMER]
delay [ONDLY]/1 shot [1SHOT]
Timer time
1 to 5000 ms (Default: 1)
[TIME]
Voltage or
current
Voltage output
*
/Current output
[V OR C]
OFF [OFF]/TASK1 [TASK1]
Output object
TASK2 [TASK2]/TASK3 [TASK3]/
[OUTPUT]
TASK4 [TASK4]
Focus
Off [OFF]
*
/On [ON]
[FOCUS]
Current: 4/5/6/7/8/9/10/11/12/13/14/
Focus
15/16/17/18/19/20 mA
Current/
(Default: 4 mA)
Voltage 1
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
[CUR1]/
1/0/1/2/3/4/5/6/7/8/9/10 V
[VOLT1]
(Default: -10 V)
Measurem
-999.999999 to 999.999999 mm
ent value 1
(Default: -0.5)
[MEAS1]
Current: 4/5/6/7/8/9/10/11/12/13/14/
Focus
15/16/17/18/19/20 mA
Current/
(Default: 20 mA)
Voltage 2
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
[CUR2]/
1/0/1/2/3/4/5/6/7/8/9/10 V
[VOLT2]
(Default: 10 V)
Measurem
-999.999999 to 999.999999 mm
ent value 2
(Default: 0.5)
[MEAS2]
Calibration
Off
*
/On
[CALIB]
Current: 4/5/6/7/8/9/10/11/12/13/14/
Calibration
15/16/17/18/19/20 mA
Current/
(Default: 4 mA)
Voltage 1
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
[CUR1]/
1/0/1/2/3/4/5/6/7/8/9/10 V
[VOLT1]
(Default: -10 V)
Calibration
value 1
-999 to 999 (Default: 0)
[ADJ1]
Current: 4/5/6/7/8/9/10/11/12/13/14/
Calibration
15/16/17/18/19/20 mA
Current/
(Default: 20 mA)
Voltage 2
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
[CUR2]/
1/0/1/2/3/4/5/6/7/8/9/10 V
[VOLT2]
(Default: 10 V)
Calibration
value 2
-999 to 999 (Default: 0)
[ADJ2]
Pages
*
p.4-28
p.4-30
*
/TASK2 [TASK2]/
p.4-18
p.4-18
p.4-19
p.4-12
*
/
p.4-13
p.4-14
p.4-16

Advertisement