Chapter 3
Setting the Output Conditions
Setting [Display]
Off [OFF]
● Hold clearing conditions
The values held can be cleared by the following operation.
• The mode is switched between the FUN mode and RUN mode.
• Hold reset is entered. (
• Hold reset command is entered.
Example: Peak
Clamp level
Example: Auto peak
Clamp level
If an abnormal measured value is obtained during sampling, the hold value is cleared. To not clear
the hold value even if an abnormal measured value is obtained, set "Keep" as the non-
measurement setting.
ZW
3-30
User's Manual
The hold function is not used. The measured value is always
output.
key)
The mode is switched between
the FUN mode and RUN mode
FUN mode
RUN mode
Hold value is reset (key input/command).
Sampling
The mode is switched between
the FUN mode and RUN mode
FUN mode
RUN mode
Hold value is reset (key input/command).
Description
Sampling
Hold value
Measurement
value
Hold value
Measurement
value
When the hold value is reset,
the clamp level is output until
the next hold value is applied.