Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform an enhanced transmission response calibration.
When you perform an enhanced transmission response calibration, the
analyzer performs correction at the selected number of data points across
the selected frequency band. Interpolation recalculates the error correction
array for reduced frequency spans. If the frequency span is increased the
calibration is invalidated, and the default response calibration is automatically
restored.
Chapter 6 provides detail about when additional calibration is necessary, and
information about other calibrations available for transmission measurements.
If you wish to calibrate your instrument for a transmission response
measurement, perform the following steps:
2. The instrument prompts you to connect four standards-open, short, load,
and through cable-as shown below.
N E T W O R K A N A L Y Z E R
OPEN
SHORT
Open, Short, load Connections
3. Press Measure Standard after connecting each standard.
Tel:0755-83589391
Response
L O A D
Fax:0755-83539691
Making Measurements
Measuring Transmission Response
R F O U T
Through Cable Connection
Website: www.Sengt.c
RF
IN
3-19