HP 8711C User Manual page 242

Rf network analyzers
Table of Contents

Advertisement

深圳市盛腾仪器仪表有限公司
Short defines the standard type of a short for calibrating
Short
reflection measurements.
Load
Load defines the standard type of a load used for calibrating
reflection measurements.
Through defines the standard type as a transmission line of
Through
For all four standard types the
must be set.
N O T E
When creating a cal kit file, if a standard is not defined, the currently defined values for that standard
will be retained. The best practice is to define all of your standards and characteristics when loading
in a new cal kit.
(usually either 50 or 75 ohms).
Delay
Delay is equivalent to a uniform length of transmission
line between the standard being dehned and the actual
measurement plane. The DELAY is entered as the one-way
travel time from the measurement plane to the standard in
seconds. Delay can be determined from the precise physical
length of the standard, multiplied by the velocity factor
(reciprocal of the dielectric constant).
Loss is used to specify energy loss, due to skin effect, along a
Loss
one-way length of coaxial delay. The value of loss is entered
as ohms/nanosecond (or Gigohmskecond) at 1 GHz. For
many applications, the loss value can be set to zero without
noticeable degradation.
Tel:0755-83589391
Calibrating for Increased Measurement
Choose an Appropriate Calibration Method
(characteristic impedance), Delay, and Loss
Fax:0755-83539691
Accuracy
6-21