HP 8711C User Manual page 141

Rf network analyzers
Table of Contents

Advertisement

Delta Amplitude
深圳市盛腾仪器仪表有限公司
This marker limit test allows you to set marker 1 as an amplitude reference
against which marker 2 is limit tested.
1.
This limit test requires that you Erst use marker 1 to determine the
reference amplitude: Press (j-1 2: and then use the front panel
knob or the Q) @J keys to place marker 1 at the desired place on the
measurement trace. The amplitude of marker 1 at this point becomes the
delta reference for this marker limit test.
2.
Press @i'GGY] Limit MEII~ Mkr Limits
Use the front panel knob or the m (IJ keys to select Delta Amp1 in
3.
the marker limit test table. Turn this limit function on by pressing the
table changes to 'on. '
4.
Press Edit Limit
limit using the front panel keypad and terminating the entry with the
k e y .
5.
Press Edit Limit (Mid&s)
using the front panel keypad and terminating the entry with the [ENTER]
key.
Tel:0755-83589391
Using Instrument Functions
Fax:0755-83539691
Using limit Testing
Website: www.Sengt.c