521 Transmlsslon Coefflclent
深圳市盛腾仪器仪表有限公司
The following graphs show typical measurement uncertainty. The
assumptions made to generate these curves were:
For transmission uncertainty, the DUT is assumed to be well-matched.
For reflection uncertainty, the DUT is assumed to be a one port device. (In
other words, load match errors produced when measuring transmission are
not taken into account.)
Power = 0 dBm for reflection measurements
Power = -20 dBm for transmission measurements
For transmission measurements, an enhanced response calibration was
performed.
For reflection measurements, a one port calibration was performed.
HP 87 11 C Uncertainty Curves
III
II
I
I
I
I
I
Tel:0755-83589391
Specifications and Characteristics
Instrument Specifications and Characteristics
Tast Port Power = 0 dBm
0
Fax:0755-83539691
I
I
I
I
10-11
Website: www.Sengt.c