Calibrating far Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
Normalization is the simplest type of calibration. The analyzer stores data
into memory and divides subsequent measurements by the stored data to
remove frequency response errors.
Follow these general steps when performing a normalization calibration:
1.
2.
3.
4.
The normalization cal will be invalidated if any frequency settings are
C A U T I O N
changed after calibration. There is no correction interpolation with a
normalization cal .
If you decide to perform any other type of calibration, be sure to turn off
C A U T I O N
normalization Iirst by pressing (KKW] l?ata . Note that the "/M" indicator
disappears when normalization is turned off.
6-10
深圳市盛腾仪器仪表有限公司
Setup the analyzer for your measurement:
Select the type of measurement
Enter operating parameters other than the default
Connect your equipment as you would for an actual measurement, but
omit the DUT.
Notice that the top of the display on the analyzer now shows the type
of measurement followed by "/M" to indicate that the displayed data is
actually the measurement divided by memory.