If averaging has been turned on, (it is off in the preset condition) reducing
the averaging factor (or turning it off altogether) will increase the analyzer's
measurement speed. Averaging requires multiple sweeps which increases
measurement time. Turning off averaging and using a narrower system
bandwidth may produce faster results.
the value displayed on the analyzer screen and press Cm].
To reduce the number of measurement points, press (jj)
numeric keypad to enter the reduced number.
Generally, as the number of points is decreased, so is the sweep time.
However, other factors will affect the sweep time such as:
using frequency bands that contain very low frequencies (below
l
approximately 20 MHz)
the number of band crossing points encountered in a sweep (at
l
approximately 1900 MHz, 2310 MHz, and 2620 MHz) (HP 8713C only)
if the MHz/msec rate is above the maximum rate at which the source can
l
be swept
The following graph shows an example of the relationship between the
number of points, frequency span, and sweep time. This graph was created
with data from a setup on an HP 8713C using a center frequency of
1500 MHz, and a system bandwidth setting of medium.