Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
How to Download
Standards Definitions
1. Insert the disk with the ca.I standard definitions into the anaIyzer's built-in
6-18
深圳市盛腾仪器仪表有限公司
N O T E
must
Calibration kit definitions
disk drive.
Press
and note the first word of the second line on the
)
RECALL
screen. If it is not "INT:\" (for internal disk), press S&W% IX&
Disk Prior
Highlight the fle (CALKIT, in the case of the Example Programs disk).
State and wait for the prompt:
Loaded cal kit information for 4 calibration standards
Recall of cal kit from CALKIT complete
Cal
Kit UWW Defined.
N O T E
Cal kit coefficients are displayed in the cal kit block of the operating parameters screen; press
(
) Operating
S Y S T E M O P T I O N S
Tel:0755-83589391
be in DOS format. LIF format is not supported for cal kit definitions.
.
Fax:0755-83539691
Website: www.Sengt.c