HP 8711C User Manual page 457

Rf network analyzers
Table of Contents

Advertisement

Specifications and Characteristics
Instrument Specifications and Characteristics
Determining Test Port
The maximum test port power output of your analyzer depends upon the
Power
options that are installed. If you have a standard instrument with no options
installed, then the values in the table preceding this note apply to your
analyzer. Otherwise, use the following table to determine your instrument's
maximum test port power:
For each option installed, subtract the indicated amount from the maximum
and minimum powers stated in the standard tables. For example, if you have
an HP 8713C with Options 1EC and 1DB installed, you would subtract a total
of 7 dB from the standard values found in the tables to get a final correct
maximum output of 3 dB for your analyzer, and - 12 dB for the minimum
output power. However, the minimum output power for any analyzer with
Option 1El (attenuator) is -60 dBm, regardless of other options installed.
If you are not sure which options (if any) are installed in your analyzer, press
[
SYSTEM OPTIONS
installed.
深圳市盛腾仪器仪表有限公司
[Attenuator]
175 ohm]
]
Source Harmonics (measured at +7 dBm*)
Frequency
Tel:0755-83589391
Oetion
subtract 1 dB
subtract 3 d6
subtract 2 dB
HP 871X
< - 2 0 dBc
< - 3 0 dBc
Fax:0755-83539691
subtract 1 dB
subtract 3 dB
subtract 4 dB
HP 871X
< - 3 0 dBc
< - 3 0 dBc
Website: www.Sengt.c