To Assemble the EI XTR Source

Materials needed
• Gloves, clean, lint-free
• Hex ball driver, 1.5 mm (8710-1570)
• Hex ball driver, 2.0 mm (8710-1804)
• Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view
page 189 while using this procedure.
1 Slide the ceramic washer into the source body.
2 Insert the extractor lens into the source body, flat side first.
3 Insert the entrance lens and ion focus lens into the insulator in the order
4 Slide the insulator containing the ion focus and entrance lens into the
5 Install the two gold plated setscrews that holds the lenses in place.
6 Assemble the repeller assembly.
Do not overtighten the repeller nut or the ceramic repeller insulators will break when
C A U T I O N
the source heats up. The nut should only be finger tight.
188
Large (8650-0030)
Small (8650-0029)
shown in
Figure 60
on page 189.
source body, with the ion focus lens against the extractor lens.
a Install the repeller block insert into the source heater block assembly.
b Install the repeller insulators into the source heater block assembly and
repeller block insert.
c Install the repeller through the repeller insulators, then put the flat
washer followed by the belleville spring washer onto the end of the
repeller shaft and secure finger tight with the repeller nut.
Figure 60
and EI XTR parts list