To Clean the CI Source

Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Solvents
• Ultrasonic bath
Preparation
1 Disassemble the CI source. (See
2 Collect the following parts to be cleaned for a CI source: (See
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
5977B Series MSD Operating Manual
Large (8650-0030)
Small (8650-0029)
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
page 214.)
page 218.)
Repeller
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
"To Disassemble the CI Source"
CI Maintenance
7
on
Figure 71
on
217