To Replace the Electron Multiplier Horn

The replacement EM horn part number for this Series 2 detector is stamped
on the front face of the detector. You can determine which series detector you
have without having to directly check the detector. The detector series is
displayed as "Triple Axis Series 2" in the detector tab of manual tune, in the
detector section on the second page of the tune report and in the pumpdown
window.
Materials needed
• Electron multiplier horn (Series 2 Detector G7002-80103)
• Gloves, clean, lint-free
• Pliers, long-nose (8710-1094)
Procedure
1 Vent the MS.
The analyzer, GC/MS interface, and other components in the analyzer chamber
WA R N I N G
operate at very high temperatures. Do not touch any part until you are sure it is cool.
Always wear clean gloves to prevent contamination when working in the analyzer
C A U T I O N
chamber.
2 Open the analyzer chamber. (See
3 Open the retaining clip. (See
4 Slide the blue signal wire from the connector in the sideplate.
5 Remove the electron multiplier horn.
6 Hold the new horn with blue signal wire end down, and attach the signal
194
Large (8650-0030)
Small (8650-0029)
(See"To Vent the MSD"
page 152.)
up and swing the clip away from the electron multiplier horn.
wire to the connector in the sideplate.
on page 113.)
"To Open the Analyzer Chamber"
Figure 63
on page 195.) Lift the arm of the clip
5977B Series MSD Operating Manual