General Maintenance
Figure 51
2 Insert the extractor ceramic insulator into the source body.
The ceramic insulator must be positioned flat against the source body when inserting
C A U T I O N
the lens stack in the next step.
3 Insert the lens stack encased in the insulator into the source body. (See
4 Using the T6 Torx driver, install and secure the gold plated setscrew and
164
Assembled EI HES
Figure 53
on page 166.) The engraved numbers face the full opening
extending to the end of the source body. Verify that the ceramic is sitting
flush at the end of the source body.
lock ring lens insulator that holds the lens assembly in place.
Align as shown
5977B Series MSD Operating Manual