To Install the CI Source

Electrostatic discharges to analyzer components are conducted to the side board
C A U T I O N
where they can damage sensitive components. Wear a grounded antistatic wrist strap 
and take other antistatic precautions before you open the analyzer chamber.
Procedure
1 Vent the MS. (See
2 Open the analyzer chamber. (See
3 Slide the CI source into the radiator. (See
4 Install the thumbscrews. (See
5 Connect the wiring to the CI source. (See
Ion source
Thumbscrews
Source radiator
Figure 74
Installing the CI source
5977B Series MSD Operating Manual
"To Vent the MSD"
page 152.)
Model Wiring from the CI Source"
on page 113.)
"To Open the Analyzer Chamber"
Figure
74.)
Figure
74.)
"To Connect/Disconnect non-HES
on page 212.)
CI Maintenance