General Maintenance
4 Abrasively clean the surfaces that contact the sample or ion beam.
5 Rinse away all abrasive residue with reagent-grade methanol.
6 Separate the parts that were abrasively cleaned from the parts that were
Always wear clean gloves to prevent contamination when working in the analyzer
C A U T I O N
chamber.
7 Ultrasonically clean the parts (each group separately) for 15 minutes in
All of these solvents are hazardous. Work in a fume hood, and take all appropriate
WA R N I N G
precautions.
8 Place the parts in a clean beaker. Loosely cover the beaker with clean
9 Dry the cleaned parts in an oven at 100 °C for 5–6 minutes.
162
Use an abrasive slurry of alumina powder and reagent-grade methanol on
a cotton swab. Use enough force to remove all discolorations. Polishing the
parts is not necessary; small scratches will not harm performance. Also,
abrasively clean the discolorations where electrons from the filaments
enter the source body.
Ensure all abrasive residue is rinsed away before ultrasonic cleaning. If
the methanol becomes cloudy or contains visible particles, rinse again.
not abrasively cleaned.
each of the following solvents:
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Methanol (reagent-grade)
aluminum foil (dull side down).
5977B Series MSD Operating Manual