5
Operating in CI Mode
Table 15
MFC
(%)
10
15
20
25
30
35
40
Familiarize yourself with the measurements on your system under operating
conditions and watch for changes that may indicate a vacuum or gas flow
problem. Measurements will vary by as much as 30% from one MSD and gauge
controller to the next.
142
MFC settings and typical pressure readings
Pressure (Torr)
Methane
EI/PCI/NCI MSD
(Turbo pump)
10
–5
5.5
×
10
–5
8.0
×
10
–4
1.0
×
10
–4
1.2
×
1
–4
1.5
×
10
–4
2.0
×
10
–4
2.5
×
Ammonia
EI/PCI/NCI MSD
(Turbo pump)
–5
5.0
×
10
–5
7.0
×
10
–5
8.5
×
10
–4
1.0
×
10
–4
1.2
×
10
–4
1.5
×
10
–4
2.0
×
10
5977B Series MSD Operating Manual