To Set the JetClean Parameters for the Clean Only Mode

Procedure:
1 In Instrument Control view, click the MS Parameters icon.
2 In the Single Quadrupole MS Method Editor, select the JetClean tab. (See
3 From the Operation dropdown menu, select Clean Only.
Figure 34
4 Enter the parameters supplied by your method developer.
5 Save the method.
5977B Series MSD Operating Manual
Figure
34.)
JetClean set to Clean Only mode
Operating in EI Mode
4