Agilent Technologies 5062-8218 Installation Note

Time gated spectrum analyzer capability installation kit

Advertisement

Quick Links

Installation Note
Time Gated Spectrum Analyzer
Capability Installation Kit, 5062-8218
Part Number 5967-8554 Supersedes: 5962-5078
Printed in USA September 2000

Advertisement

loading

Summary of Contents for Agilent Technologies 5062-8218

  • Page 1 Installation Note Time Gated Spectrum Analyzer Capability Installation Kit, 5062-8218 Part Number 5967-8554 Supersedes: 5962-5078 Printed in USA September 2000...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
  • Page 3: Installation Kit Parts List

    To Be Performed By: ..... . Agilent Technologies Service Center Only Estimated Installation Time: ....
  • Page 4: Tools Required

    The firmware date is displayed on the analyzer screen. The firmware is displayed in a day.month.year format. See Contacting Agilent Technologies for details. WARNING Before disassembling the instrument, turn the power switch OFF and unplug the analyzer.
  • Page 5 Figure 1 Removing The instrument Cover 5. Pull the instrument cover off toward the rear of the instrument. 6. If there is a dual I/O option on the back of the instrument it must be removed. If the instrument does not have the dual I/O option, then go to step 9. 7.
  • Page 6 If J501 and J500 do not exist on the CPU, you will need to order a CPU Board Kit. Refer to Contacting Agilent Technologies section of this document. 15. Being careful not to pinch or damage any of the cables, reattach the rear panel assembly to the chassis of the analyzer.
  • Page 7 Figure 3 Attaching Cables to Board Assembly Installation Note 5967-8554...
  • Page 8: Contacting Agilent Technologies

    22. Tighten the four rear-feet screws with a 4-mm hex wrench. 23. Replace the four screws and washers attaching the instrument cover assembly to the rear frame. Contacting Agilent Technologies 1. If the CPU board does not have J501 and J500, order the 08590-60361 CPU kit for a compatible CPU board.