These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
If insulators are dirty, clean them with a cotton swab dampened with reagent-grade
C A U T I O N
methanol. If that does not clean the insulators, replace them. Do not abrasively or
ultrasonically clean the insulators.
Figure 50
The filaments, source heater assembly, and insulators cannot be cleaned
C A U T I O N
ultrasonically. Replace these components if major contamination occurs.
3 If the contamination is serious, such as an oil backflow into the analyzer,
.
Do not use the abrasive slurry on the source mount bushings.
C A U T I O N
5977B Series MSD Operating Manual
Source body
Lens
1
EI HES parts to be cleaned
seriously consider replacing the contaminated parts
Source mount
Repeller
with bushings
2
3
General Maintenance
Filament block
4
5
6
161