6
General Maintenance

To Connect/Disconnect Wiring to the EI HES

Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Pliers, long-nose (8710-1094)
• Tweezers (8710-2460)
Procedure
1 Use tweezers or needle nose pliers to connect/disconnect the ceramic board
Figure 48
Wiring between the ceramic source board and the EI HES
156
wire leads (red, white, black, and gray) at the source connectors. (See
Figure
48.)
Repeller
Filament common
Filament 1
Filament 2
Ceramic source board
5977B Series MSD Operating Manual