To Install the EI XTR, SS, or Inert Source

Materials needed
• Gloves, clean, lint-free
• Pliers, long-nose (8710-1094)
Procedure
1 Slide the ion source into the source radiator. (See
Ion source
Ceramic source board
Thumbscrews
Extractor wire
Source heater and temperature
sensor wires
Source radiator
Figure 62
Installing the EI source
2 Insert and hand tighten the source thumbscrews. Do not overtighten the
3 Connect the ion source wires. (See
4 Close the analyzer chamber. (See
5977B Series MSD Operating Manual
Large (8650-0030)
Small (8650-0029)
thumbscrews. (See
Figure
EI XTR, SS, and Inert Sources"
page 196.)
62.)
"To Connect/Disconnect Wiring from the
on page 173.)
"To Close the Analyzer Chamber"
General Maintenance
Figure
62.)
on
6
193