To Connect/Disconnect Wiring from the EI XTR, SS, and Inert
Sources
Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Pliers, long-nose (8710-1094)
• Tweezers (8710-2460)
Procedure
1 Use tweezers or needle nose pliers to connect/disconnect the ceramic
2 Use tweezers or needle nose pliers to connect/disconnect the source heater
Entrance lens
Ion focus lens
Filament 1
Repeller
Filament 2
Extractor lens
Source heater
Source RTD
Figure 55
Wiring for the EI XTR Source
5977B Series MSD Operating Manual
source board wire leads at the source connectors. (See
wire leads at the ceramic source board.
Ceramic source board
General Maintenance
Figure
55.)
6