CI Maintenance

To Remove the CI Source Radiator

Materials needed
• Gloves, clean, lint-free
• Tweezers (8710-2460)
• T10 Torx screwdriver (5182-3466)
Procedure
The analyzers, GC/MS interface, and other components in the analyzer chamber
WA R N I N G
operate at very high temperatures. Do not touch any part until you are sure it is cool.
Always wear clean gloves to prevent contamination when working in the analyzer
C A U T I O N
chamber.
Make sure you use an antistatic wrist strap and take other antistatic precautions
C A U T I O N
before touching analyzer components.
When disconnecting leads, pull on the connectors, not on the wires.
C A U T I O N
206
Large (8650-0030)
Small (8650-0029)
5977B Series MSD Operating Manual