Agilent Technologies 5977B Series Operating Manual page 122

Msd
Hide thumbs Also See for 5977B Series:
Table of Contents

Advertisement

5
Operating in CI Mode
Table 10
Reagent gas settings
Reagent gas
Ion polarity
Emission
Electron energy
Filament
Repeller
Ion focus
Entrance lens offset
EM volts
Shutoff valve
Gas select
Suggested flow
Source temp
Quad temp
Interface temp
Autotune
* N/A Not Available
122
Methane
Positive
Negative
150 A
50 A
150 eV
150 eV
1
1
3 V
3 V
130 V
130 V
20 V
20 V
1200
1400
Open
Open
A
A
20%
40%
250 °C
150 °C
150 °C
150 °C
280 °C
280 °C
Yes
Yes
Isobutane
Positive
Negative
150 A
50 A
150 eV
150 eV
1
1
3 V
3 V
130 V
130 V
20 V
20 V
1200
1400
Open
Open
B
B
20%
40%
250 °C
150 °C
150 °C
150 °C
280 °C
280 °C
No
Yes
Ammonia
Positive
Negative
150 A
50 A
150 eV
150 eV
1
1
3 V
3 V
130 V
130 V
20 V
20 V
1200
1400
Open
Open
B
B
20%
40%
250 °C
150 °C
150 °C
150 °C
280 °C
280 °C
No
Yes
5977B Series MSD Operating Manual
EI
*
N/A
35 A
70 eV
1 or 2
30 V
90 V
25 V
1300
Closed
None
N/A
230 °C
150 °C
280 °C
Yes