To Replace a Filament in an EI XTR, SS, or Inert Source
Materials needed
• Filament assembly (G7005-60061)
• Gloves, clean, lint-free
• Hex ball driver, 1.5-mm (8710-1570)
Procedure
1 Vent the MSD. (See
The analyzer operates at high temperatures. Do not touch any part until you are sure
WA R N I N G
it is cool.
2 Open the analyzer chamber. (See
3 Remove the ion source. (See
4 Remove the gold plated screw and washer for the filaments. (See
5977B Series MSD Operating Manual
•
Large (8650-0030)
•
Small (8650-0029)
"Venting the MSD"
page 152.)
page 172.)
on page 192.)
on page 91.)
"To Open the Analyzer Chamber"
"To Remove the EI XTR, SS, or Inert Source"