Agilent Technologies 5977B Series Operating Manual page 165

Msd
Hide thumbs Also See for 5977B Series:
Table of Contents

Advertisement

Figure 52
5 Place the heater/sensor assembly over the guide pins on the source body
6 Place the repeller onto the heater/sensor assembly with the flat side of the
7 Place the source mount over the repeller.
8 Finger tighten the two gold plated screws using a T6 Torx screwdriver to
Do not overtighten the screws in the source body. Damage may occur to the repeller.
C A U T I O N
9 Thread the finger grip into the filament block.
10 On the side of the filament block opposite the finger grip, orient the dual
11 Inset the filament block into the source holder, and use a T6 Torx
5977B Series MSD Operating Manual
Secure the lens screw and lock ring insulator
with the four electrical pins oriented down the flat side of the source body.
repeller circumference aligned with the interface socket in the source body.
secure the source mount to the source body.
filament ceramic holder so that it lines up with the flat of the filament
block. Fully insert the three filament leads into the filament block.
screwdriver to secure it to the holder with the gold plated screw.
General Maintenance
6
165