To Connect/Disconnect EI HES Model Wiring to the CI Source

Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Pliers, long-nose (8710-1094)
• Tweezers (8710-2460)
Procedure
1 Use pliers to connect/disconnect the ceramic board wire leads (red, white,
2 Use pliers to connect/disconnect the source heater wire leads at the
Entrance lens
Ion focus
Filament 1
Filament 1
Repeller
Dummy filament
Source heater
Source RTD
Figure 69
Wiring between the ceramic source board and the source
5977B Series MSD Operating Manual
black, and gray) at the source connectors. (See
ceramic source board. (See
Ceramic source board
Figure
Figure
69.)
CI Maintenance
69.)
213
7