Compression bolts
Click and run connectors
Column clamps
Retainer ring
SmartID key
Figure 12
8
9
10 Remove and store the column for later use in accordance with the column
5977B Series MSD Operating Manual
9000 GC column and related parts
Unplug the column SmartID key from the lower USB port.
Using the Intuvo torque driver, remove the two compression bolts that seal
the column's click and run connectors to the bus and store for later use.
manufacturer's recommendations.
Intuvo 9000 GC Column Maintenance