To Remove the CI Source Filament

Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Hex ball driver, 1.5 mm (8710-1570)
• Tweezers (8710-2460)
Procedure
1 Vent the MS. (See
Always wear clean gloves to prevent contamination when working in the analyzer chamber.
C A U T I O N
The analyzer, GC/MS interface, and other components in the analyzer chamber
WA R N I N G
operate at very high temperatures. Do not touch any part until you are sure it is cool.
2 Open the analyzer chamber. (See
3 Remove the CI source. (See
4 Remove the screw holding the filament to the CI source body. (See
5 Slide the filament off the CI source assembly. (See
Figure 73
5977B Series MSD Operating Manual
"To Vent the MSD"
Figure
73.)
Screw
Filament
Changing the CI source filament
on page 113.)
"To Open the Analyzer Chamber"
"To Remove the CI Source"
CI Maintenance
on page 152.)
on page 211.)
Figure
73.)
CI source body
7
223