CI Maintenance

To Install a CI Source Filament

Materials needed
• Filament assembly, 2-pk, CI (G7005-60072)
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Tweezers (8710-2460)
Procedure
1 Remove the old filament. (See
2 Place the new filament into its position on the ion source body. (See
3 Secure the filament to the ion source body with the screw. (See
4 After installing the filament, verify that it is not grounded to source body.
5 Reinstall the CI source. (See
6 Pumpdown the MSD. (See
7 Autotune the MSD.