15 Gently check the alignment of the ion source and the interface tip seal.
Forcing the analyzer door closed if these parts are misaligned will damage the tip seal
C A U T I O N
or the interface or the ion source, or will keep the sideplate from sealing.
16 You can align the ion source and interface tip seal by wiggling the side plate
17 Close the analyzer chamber door. (See
46
When the ion source is aligned correctly, the analyzer chamber can be
closed all the way with no resistance except the spring tension from the
interface tip seal.
on its hinge. If the door still will not close, contact your Agilent
Technologies service representative.
page 196.)
"To Close the Analyzer Chamber"