6
General Maintenance
To Install the EI HES Filament
Materials needed
• Filament assembly, high efficiency dual filament (G7001-60061)
• Hex ball driver, 1.5 mm (8710-1570)
• Gloves, clean, lint-free
• T6 Torx driver, (8710-2548)
Procedure
1 Remove the old filament. (See
2 Insert the 3 pins on the dual filament through the rear of the filament block.
3 Place the filament block in the source mount.
4 Use a T6 Torx screwdriver to tighten the screw securing the filament block
5 Reinstall the ion source. (See
6 Close the analyzer chamber. (See
7 Pumpdown the MSD. (See
8 Autotune the MSD.
170
•
Large (8650-0030)
•
Small (8650-0029)
page 168.)
(See
Figure 54
on page 169.)
to the source mount.
page 196.)
"To Remove the EI HES filaments"
"To Install the EI HES"
"To Close the Analyzer Chamber"
"To Pumpdown the MSD"
5977B Series MSD Operating Manual
on
on page 171.)
on
on page 116.)