3
Intuvo 9000 GC Column Maintenance
11 Push the MS tail heater assembly back a few millimeters. A magnet will
Ferrule release hole
Transfer line
Bus click and run connection
Figure 16
12 Remove the 9000 GC/MS Tail from the transfer line and bus. 
Exit in ion source
Figure 17
13 Replace the gasket. (See
14 Carefully slide the 9000 GC/MS Tail into the GC/MS interface, and gently
15 Check that the 9000 GC/MS Tail click and run fitting sits flat against the
72
hold the heater assembly away from the 9000 GC/MS Tail.
MS tail heater assembly clamp open with 9000 GC/MS Tail removed
If the ferrule becomes stuck, press a pointed object, like a paper clip, into
the ferrule release hole at the end of the transfer line.
9000 GC/MS Tail removed from GC/MS Interface
page 65.)
place the click and run connection into the bus connection.
gasket in the bus connection.
Ferrule and nut
Click and run connection
"To Replace an Intuvo 9000 GC Gasket"