The GC/MSD Interface for a 7890B Series GC

The GC/MSD interface is a heated conduit into the MSD for the capillary
column. (See
chamber, with an O-ring seal. It has a protective cover which should be left in
place.
One end of the interface passes through the side of the GC and extends into
the oven. This end is threaded to allow connection of the column with a nut
and ferrule. The other end of the interface fits into the ion source. The end of
the capillary column extends slightly past the end of the column guide tube
and into the ionization chamber.
Ionization
chamber
Ion source
Analyzer
chamber
Heater/Sensor
assembly
Figure 10
GC/MSD interface for a 7890B GC with EI XTR, HES, and CI sources
5977B Series MSD Operating Manual
Figure
10.) It is bolted onto the right side of the analyzer
Tip seal
assembly
Installing 7890 GC Columns
Heater sleeve
Insulation
Column
guide tube
Column
MSD
GC oven
Optional entry for CI reagent
gas or JetClean system
hydrogen
2
55