To Set The Masshunter 7890 Gc/Msd Interface Temperature - Agilent Technologies 5977B Series Operating Manual

Msd
Hide thumbs Also See for 5977B Series:
Table of Contents

Advertisement

4
Operating in EI Mode

To Set the MassHunter 7890 GC/MSD Interface Temperature

Procedure
1 From Instrument Control view select Instrument>GC Edit Parameters.
2 Click the Aux Heaters icon to edit the interface temperature. (See
Figure 30
GC edit parameters window
3 Select On to turn on the heater and type the setpoint in the Value °C column.
Ensure that the carrier gas is turned on and the column has been purged of air before
C A U T I O N
heating the GC/MS interface or the GC oven.
When setting the GC/MS interface temperature, never exceed the maximum for your
column.
4 Click Apply to download setpoints, or click OK to download setpoints and
5 To make the new settings part of the method, select Save from the Method
100
The typical setpoint is 280 °C.
close the window.
menu.
5977B Series MSD Operating Manual
Figure
30.)