6
General Maintenance
To Clean the EI HES
Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (393706201)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Solvents
• Ultrasonic bath
Procedure
1 Disassemble the EI HES. (See
2 Collect the following parts to be cleaned: (See
160
•
Large (8650-0030)
•
Small (8650-0029)
•
Methanol (reagent-grade)
•
Methylene chloride (reagent-grade)
•
Acetone (reagent-grade)
•
Filament mount
•
Source mount (does not get abrasively or ultrasonically cleaned)
•
Repeller
•
Source body
•
Extractor lens (5)
•
Post extractor lens 1 (4)
•
Post extractor lens 2 (3)
•
Ion focus lens (2)
•
Entrance lens (1)
"To Disassemble the EI HES"
Figure 50
5977B Series MSD Operating Manual
on page 157.)
on page 161.)