GC inlet cover removed
Compression bolt access plate
Figure 13
7
8
9
5977B Series MSD Operating Manual
From MassHunter Data Acquisition, Vent the MS (See
on page 113.)
Wait until the GC reaches the ready state indicating the components are
cooled below 40 °C and the instrument is vented before continuing the
steps in this procedure.
If you are using hydrogen or other flammable gas as a carrier gas or for the
JetClean system, close the manual gas supply valve to the instrument
before turning off the MS power.
If installed, remove the ALS injector from the inlet.
Remove the inlet cover.
GC inlet cover and compression bolt access plate
Open the GC front door.
Open the bus door and remove it by lifting the door vertically off its hinge
pins.
Pull the compression bolt access plate out to allow the torque driver to
access the Guard chip compression bolt.
Intuvo 9000 GC Column Maintenance
"To Vent the MSD"
3
67