Agilent Technologies 5977B Series Operating Manual page 219

Msd
Hide thumbs Also See for 5977B Series:
Table of Contents

Advertisement

5 Ultrasonically clean the parts (each group separately) for 15 minutes. For
All of these solvents are hazardous. Work in a fume hood and take all appropriate
WA R N I N G
precautions.
Al
6 Place the parts in a clean beaker. Loosely cover the beaker with clean
7 Dry the cleaned parts in an oven at 100 °C for five to six minutes.
Let the parts cool before you handle them.
WA R N I N G
Let
Take care to avoid contaminating cleaned and dried parts. Put on new, clean gloves before
N O T E
handling the parts. Do not set the cleaned parts on a dirty surface. Set them only on clean,
lint-free cloths.
5977B Series MSD Operating Manual
dirty parts, use all three solvents in the order shown, cleaning 15 minutes
with each of the following solvents:
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Methanol (reagent-grade)
For routine cleaning, cleaning with methanol is sufficient.
aluminum foil (dull side down).
CI Maintenance
7
219

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5975 msd5977 msd