To Connect/Disconnect Non-Hes Model Wiring From The Ci Source - Agilent Technologies 5977B Series Operating Manual

Msd
Hide thumbs Also See for 5977B Series:
Table of Contents

Advertisement

7
CI Maintenance

To Connect/Disconnect non-HES Model Wiring from the CI Source

Materials needed
• Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
• Pliers, long-nose (8710-1094)
• Tweezers (8710-2460)
Procedure
1 Use pliers to connect/disconnect the ceramic board wire leads at the source
2 Use pliers to connect/disconnect the source heater wire leads at the
Entrance lens
Ion focus
Filament 1
Repeller
Dummy filament
Source heater
Source RTD
Figure 68
Wiring between the ceramic source board and the source
212
connectors. (See
Figure
ceramic source board. (See
Ceramic source board
68.)
Figure
68.)
5977B Series MSD Operating Manual