Advanced Graphic Data Manager
"Diagnosis/
Simulation" menu
items
LCD operating time
Submenu: Hardware
Submenu: Protocols
Internal unit logs.
Boot log
Error log
Submenu: Simulation
Various functions/signals can be simulated for test purposes here.
Note: During the simulation mode normal recording of the measured values is interrupted and the intervention is logged in the event
log.
Outputs
Display test
Test bar code reader
Initialize modem
Initializes the modem connected (to automatically answer calls).
Note: The modem has to be connected to the serial interface of the device. Only use the RXU10-A1 modem cable (see Accessories,
Section 8). Beforehand, set the baud rate to be used to transmit the data under "Setup -> System -> Communication -> Serial
interface".
Submenu: Memory
Information on the size of the memory and how long data can be saved.
information
Internal memory
SD card
Delete internal
Deletes the entire internal measured value memory.
Note: The SD card or the USB stick is not cleared. If a release code has been set up then the memory can only be deleted after this
memory
code has been entered. When the user administration function is active, this action can only be performed by the administrator.
Calibration
Calibration must only be carried out by skilled and trained personnel.
Note: Malfunctions possible if not performed correctly! Normal recording of the measured values is interrupted while this function
is being used and the intervention is logged in the event log. Can only be changed using the service code.
Configurable parameters
(factory settings are highlighted in bold)
Indicates how long the display of the unit was in operation.
Information on the hardware components, and what slots are assigned, with
information on the software version.
Submenu: USB devices
Overview of the USB devices currently connected.
Device 1-8: information on the USB device connected.
Logs the boot procedure of the unit (for diagnosis purposes).
Note: Can only be accessed using the service code.
Only for diagnosis purposes.
Note: Can only be accessed using the service code.
Output simulation (analog/pulse output, relay).
Note: Only possible if switched on/assigned.
Test whether all pixels are being controlled or whether the colors can be easily identified
from each other.
Test the function of the bar code reader (e.g. to determine if the right character set is
being used).
Note: Only possible if a bar code reader is connected to the device.
Indicates the size of the memory in MB (or GB) and how long data can be saved until
the memory is full.
Note:
• Normal operation (i.e. standard save cycle) is presumed when calculating the
memory duration. Higher save speeds, due to alarms or many event messages, can
reduce the storage duration considerably!
• With regard to external storage media, third-party files, or the creation of screenshots,
can also reduce the duration of the memory. For this reason, read out the memory on
time or change the external storage medium.
Note: Even if the (external) memory is not yet full, it is advisable to read out the data
frequently or to back up the data on the PC.
The storage information for a USB stick is not displayed.
Commissioning and use during operation
99